Energy dependence of an epitaxial diode in standard diagnostic radiology beams

Carregando...
Imagem de Miniatura

Data

Data de publicação

2024

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

É parte de

INTERNATIONAL SYMPOSIA ON RADIATION PHYSICS, 19th
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
The overall response of a radiation-hard epitaxial diode has been previously investigated for radiation diagnostic radiology qualities (RQR-3, RQR-5, RQR-8, and RQR-10) and computed tomography qualities (RQT-8, RQT-9, and RQT-10) beams. The EPI diode has a thin n-type epitaxial layer (25-75 µm) grown on a thick (300-500 µm) Czochralski silicon substrate and a p-n junction provided by a highly doped ptype silicon layer ( 1 µm). When operating as a dosimeter in the short-circuit current mode, the dosimetric parameters of the diode regarding repeatability (< 0.3%), long-term stability (0.4%/year), angular response (< 3%, ± 5º), dose rate dependence (< 3%), and signal-to-noise ratio (≥ 1500) fully adhered to the IEC 61674 recommendations [1]. However, compliance with the energy dependence requirement (≤ 5%) was not achieved for RQR-10 and RQT-10 beams, with average energies of 60 keV upward. This unexpected dependence, experimentally manifested by variations in the charge sensitivities, might be associated with the physical phenomena underlying the photon interaction with the complex design of the EPI diode. This work aims to provide a theoretical basis for the data previously gathered with the EPI diode and validate its use as a dosimeter for low-energy photon beams. It has been accomplished through current and charge sensitivities calculations considering the diode as a thin abrupt junction supported on a highly doped Czochralski substrate, the minority carriers' diffusion lengths, and the X-ray energy spectra from a PantakSeifert 160HS Isovolt X-ray generator. The theoretical results are compared to the experimental ones, as visible in Fig.1. Despite the overall data agreement in the low-energy region, the discrepancy between the experimental and calculated values for photons of 60 keV upward remains to be investigated. Studies in this direction are underway.

Como referenciar
GONÇALVES, J.A.C.; MANGIAROTTI, A.; ANTONIO, P.L.; CALDAS, L.V.E.; BUENO, C.C. Energy dependence of an epitaxial diode in standard diagnostic radiology beams. In: PESSANHA, SOFIA (ed.); MANSO, MARTA (ed.). In: INTERNATIONAL SYMPOSIA ON RADIATION PHYSICS, 19th, September 1-5, 2024, Lisbon, Portugal. Abstract... Lisbon, Portugal: Universidade Nova de Lisboa - Faculdade de Ciências e Tecnologia, 2024. p. 165-165. Disponível em: https://repositorio.ipen.br/handle/123456789/48968. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento