Use of a standard ionization chamber at different distances in diagnostic radiology beams
Carregando...
Data
Data de publicação
2009
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
INTERNATIONAL CONGRESS OF THE IUPESM, 11th
Resumo
A secondary standard ionization chamber was
calibrated, by a primary standard laboratory, at a determined
distance. However, in the routine procedures of a calibration
laboratory, sometimes the radiation detector cannot be positioned exactly at the calibration distance, for any reason. In
this work, the response of a secondary standard ionization
chamber was tested to verify the possibility of its use at different distances than at the calibration distance, using the same
calibration factor.
Como referenciar
YOSHIZUMI, M.T.; AFONSO, L.C.; CALDAS, L.V.E. Use of a standard ionization chamber at different distances in diagnostic radiology beams. In: INTERNATIONAL CONGRESS OF THE IUPESM, 11th, September 7-12, 2009, Munique, Alemanha. Proceedings... Disponível em: http://repositorio.ipen.br/handle/123456789/16706. Acesso em: 30 Dec 2025.