Primary standardization of sup(242)Am radioactive sources

Carregando...
Imagem de Miniatura
Data
Data de publicação:
2001
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
REGIONAL CONGRESS ON RADIATION PROTECTION AND SAFETY, 5th, REGIONAL IRPA CONGRESS
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
The procedure followed by the Laboratório de Metrologia Nuclear in São Paulo, Brazil, for the standardization of 242gAm is described. The calibration system was composed of a 4π gas-flow proportional counter coupled to a pair of NaI(Tl) crystals operating in coincidence. The samples were produced by irradiating dried aliquots of 241Am with thermal and epithermal neutrons at the IEA-R1 research reactor. The efficiency tracer technique has been applied using 60Co as tracer. The beta detection efficiency was changed by external absorbers and extrapolated to unity by linear least square fitting applying covariance methodology.

Como referenciar
MAIDANA, N.L.; DIAS, M.S.; KOSKINAS, M.F. Primary standardization of sup(242)Am radioactive sources. In: REGIONAL CONGRESS ON RADIATION PROTECTION AND SAFETY, 5th, REGIONAL IRPA CONGRESS, Apr. 29 - May 4, 2001, Recife, PE. Proceedings... Disponível em: http://repositorio.ipen.br/handle/123456789/15242. Acesso em: 14 Mar 2025.