Evaluation of some dosimetric properties of a dedicated plane parallel ionization chamber for radiotherapy electron beams

dc.contributor.authorSILVA, J.O.
dc.contributor.authorNONATO, F.B.C.
dc.contributor.authorSAMPAIO, F.G.A.
dc.contributor.authorCALDAS, L.V.E.
dc.coverageNacionalpt_BR
dc.date.accessioned2018-12-10T12:42:40Z
dc.date.available2018-12-10T12:42:40Z
dc.date.issued2018pt_BR
dc.description.abstractIonization chambers are the reference detectors for electron beam dosimetry. In this paper a dedicated radiotherapy plane parallel ionization chamber manufactured with low cost materials is presented for dosimetry in electron beams. The ionization chamber tested has a sensitive volume of 0.4 cm³. Both the collecting electrode and the guard ring were painted with a homogeneous mixture of nail polish and graphite. The dedicated ionization chamber presented a linear response with electron absorbed dose within the range 0.5 to 8.0 Gy, an increase of its response with the field size increasing, an angular dependence within ±5°, as recommended by international standards, and a polarity effect of 0.78% according to the field size. Considering the results obtained, it is possible to conclude that the plane parallel ionization chamber tested in this work presents potential use for elec - tron beam dosimetry in clinical routine.pt_BR
dc.description.sponsorshipConselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)pt_BR
dc.description.sponsorshipFundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)pt_BR
dc.description.sponsorshipInstituto Nacional de Ciência e Tecnologia (INCT)pt_BR
dc.description.sponsorshipIDCNPq: 301335/2016-8pt_BR
dc.description.sponsorshipIDFAPESP: 08/57863-2pt_BR
dc.description.sponsorshipIDINCT: 573659/2008-7pt_BR
dc.format.extent1-13pt_BR
dc.identifier.citationSILVA, J.O.; NONATO, F.B.C.; SAMPAIO, F.G.A.; CALDAS, L.V.E. Evaluation of some dosimetric properties of a dedicated plane parallel ionization chamber for radiotherapy electron beams. <b>Brazilian Journal of Radiation Sciences</b>, v. 6, n. 3, p. 1-13, 2018. DOI: <a href="https://dx.doi.org/10.15392/bjrs.v6i3.547">10.15392/bjrs.v6i3.547</a>. Disponível em: http://repositorio.ipen.br/handle/123456789/29345.
dc.identifier.doi10.15392/bjrs.v6i3.547pt_BR
dc.identifier.fasciculo3pt_BR
dc.identifier.issn2319-0612pt_BR
dc.identifier.orcidaguardandopt_BR
dc.identifier.orcidhttps://orcid.org/0000-0002-7362-2455
dc.identifier.urihttp://repositorio.ipen.br/handle/123456789/29345
dc.identifier.vol6pt_BR
dc.relation.ispartofBrazilian Journal of Radiation Sciencespt_BR
dc.rightsopenAccesspt_BR
dc.subjectabsorbed radiation doses
dc.subjectdosimetry
dc.subjectelectron beams
dc.subjectionization chambers
dc.subjectperformance
dc.subjectphantoms
dc.subjectplanning
dc.subjectradiotherapy
dc.titleEvaluation of some dosimetric properties of a dedicated plane parallel ionization chamber for radiotherapy electron beamspt_BR
dc.typeArtigo de periódicopt_BR
dspace.entity.typePublication
ipen.autorLINDA V. E. CALDAS
ipen.codigoautor1495
ipen.contributor.ipenauthorLINDA V. E. CALDAS
ipen.date.recebimento18-12pt_BR
ipen.identifier.fiSem F.I.pt_BR
ipen.identifier.ipendoc25138pt_BR
ipen.type.genreArtigo
relation.isAuthorOfPublication7f46d4f4-dfd6-4485-a767-10df5b4f4f13
relation.isAuthorOfPublication.latestForDiscovery7f46d4f4-dfd6-4485-a767-10df5b4f4f13
sigepi.autor.atividadeCALDAS, L.V.E.:1495:330:Npt_BR

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
25138.pdf
Tamanho:
491.04 KB
Formato:
Adobe Portable Document Format
Descrição:

Licença do Pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:

Coleções