Influence of thermal treatments on the response of sand radiation detectors for high-dose dosimetry

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Radiation Protection Dosimetry
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
The dosimetric properties of sand from Brazilian beaches have shown to be useful for high-dose dosimetry. The thermoluminescent (TL) and electron paramagnetic resonance (EPR) techniques were utilised, and the sand samples were recently studied in relation to their main dosimetric properties. The EPR signal at g = 1.999 grows significantly in function of the absorbed dose, and the TL peaks appear at 110 and 170ºC. However, these sand samples present a post-irradiation thermal decay at room temperature, which is a problem for dosimetric procedures. In this study, sand samples have been studied in relation to different thermal treatments. Post-irradiation treatments were performed at 50ºC up to 230ºC.

Como referenciar
CALDAS, LINDA V.E.; TEIXEIRA, MARIA I.; FERRAZ, GILBERTO M. Influence of thermal treatments on the response of sand radiation detectors for high-dose dosimetry. Radiation Protection Dosimetry, v. 120, n. 1-4, p. 230-234, 2006. DOI: 10.1093/rpd/nci660. Disponível em: http://repositorio.ipen.br/handle/123456789/5437. Acesso em: 30 Dec 2025.