Evaluation of a special parallel-plate ionization chamber in standard mammography beams
Carregando...
Data
Data de publicação
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Applied Radiation and Isotopes
Resumo
The aim of this work was to verify the possibility of the application of a special parallel-plate ionization chamber,
developed at the Calibration Laboratory (IPEN), for mammography dosimetry. The homemade chamber has a
polymethyl methacrylate body, 10 cm of sensitive length, and a sensitive volume of 3.2 cm3. Its entrance window
is a foil of Mylar and its collecting electrode is made of graphite. Some quality control and characterization tests
were applied to evaluate the ionization chamber: short- and medium-term stabilities, leakage current, saturation,
ion collection efficiency, polarity effect, stabilization time, linearity of response, angular dependence, and energy
dependence. The tests were carried out using mammography standard radiation beams and a90Sr + 90Y check
source. All the results obtained were within the international recommended limits.
Como referenciar
SOARES, CLIMERIO S.; CALDAS, LINDA V.E. Evaluation of a special parallel-plate ionization chamber in standard mammography beams. Applied Radiation and Isotopes, v. 225, p. 1-5, 2025. DOI: 10.1016/j.apradiso.2025.111991. Disponível em: https://repositorio.ipen.br/handle/123456789/49341. Acesso em: 20 Mar 2026.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.