Calibration coefficients of epitaxial diodes used in diagnostic radiology and computed tomography beams

dc.contributor.authorGONCALVES, J.A.C.
dc.contributor.authorANTONIO, P.L.
dc.contributor.authorCALDAS, L.V.E.
dc.contributor.authorBUENO, C.C.
dc.coverageNacional
dc.date.accessioned2026-01-09T16:02:27Z
dc.date.available2026-01-09T16:02:27Z
dc.date.issued2024
dc.description.abstractThe assessment of the calibration coefficients of an epitaxial diode, previously characterized for diagnostic radiology (RQR qualities) and computed tomography (RQT qualities) dosimetry, is reported in this work. The diode, with an n-type epitaxial layer (50 µm) grown on a thick (300 µm) Czochralski silicon substrate, is directly connected to an electrometer Keithley 6517B in the photovoltaic mode and exposed to X-ray beams from a Pantak/Seifert generator, model Isovolt 160 HS. Under this operational condition, the dosimetric quantity is the dose rate correlated with the output current signal from the diode when exposed to radiation. The corresponding collected charge (the integral of the current signal) is proportional to the dose. The repeatability of the current signals and the dose response of the diode are investigated in several RQR and RQT beam qualities spanning from 50 kV to 150 kV. The calibration coefficients (Nk) are assessed by linking the diode readings to those from the standard ionization chambers for reference beam qualities (RQR-5 and RQT-9) following the formalism recommended by Technical Reports Series No. 457. To take into account the variations of the X-ray beams related to the correspondent reference beam qualities, the correction factors are also calculated. Despite being close to 1, they evidence the energy dependence of the EPI diode for voltages above 100 kV.
dc.description.sponsorshipFundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
dc.description.sponsorshipConselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)
dc.description.sponsorshipIDFAPESP: 18/05982-0; 22/13430-2
dc.description.sponsorshipIDCNPq: 406303/2022-3; 305142/2021-6
dc.format.extent1-10
dc.identifier.citationGONCALVES, J.A.C.; ANTONIO, P.L.; CALDAS, L.V.E.; BUENO, C.C. Calibration coefficients of epitaxial diodes used in diagnostic radiology and computed tomography beams. <b>Brazilian Journal of Radiation Sciences</b>, v. 12, n. 4A, p. 1-10, 2024. DOI: <a href="https://dx.doi.org/10.15392/2319-0612.2024.2576">10.15392/2319-0612.2024.2576</a>. Disponível em: https://repositorio.ipen.br/handle/123456789/49051.
dc.identifier.doi10.15392/2319-0612.2024.2576
dc.identifier.fasciculo4A
dc.identifier.issn2319-0612
dc.identifier.orcidhttps://orcid.org/0000-0002-8940-9544
dc.identifier.orcidhttps://orcid.org/0000-0002-7362-2455
dc.identifier.orcidhttps://orcid.org/0000-0001-7881-7254
dc.identifier.percentilfiSem Percentil F.I.
dc.identifier.percentilfiCiteScoreSem Percentil CiteScore
dc.identifier.urihttps://repositorio.ipen.br/handle/123456789/49051
dc.identifier.vol12
dc.language.isoeng
dc.relation.ispartofBrazilian Journal of Radiation Sciences
dc.rightsopenAccess
dc.sourceInternational Nuclear Atlantic Conference, May 6-10, 2024, Rio de Janeiro, RJ
dc.titleCalibration coefficients of epitaxial diodes used in diagnostic radiology and computed tomography beams
dc.title.alternativeCoeficientes de calibração de diodos epitaxiais utilizados em feixes de radiologia diagnóstica e tomografia computadorizada
dc.typeArtigo de periódico
dspace.entity.typePublication
ipen.autorJOSEMARY ANGELICA CORREA GONCALVES
ipen.autorPATRICIA DE LARA ANTONIO
ipen.autorLINDA V. E. CALDAS
ipen.autorCARMEN CECILIA BUENO TOBIAS
ipen.codigoautor924
ipen.codigoautor6585
ipen.codigoautor1495
ipen.codigoautor1592
ipen.contributor.ipenauthorJOSEMARY ANGELICA CORREA GONCALVES
ipen.contributor.ipenauthorPATRICIA DE LARA ANTONIO
ipen.contributor.ipenauthorLINDA V. E. CALDAS
ipen.contributor.ipenauthorCARMEN CECILIA BUENO TOBIAS
ipen.identifier.fiSem F.I.
ipen.identifier.fiCiteScoreSem CiteScore
ipen.identifier.ipendoc31146
ipen.type.genreArtigo
relation.isAuthorOfPublication76fdc4d1-7624-4332-a9d0-f06826000679
relation.isAuthorOfPublication5ea4ca8b-8415-4439-81d2-6fe0d6debd42
relation.isAuthorOfPublication7f46d4f4-dfd6-4485-a767-10df5b4f4f13
relation.isAuthorOfPublicationfa74399b-83a0-4f46-91e0-da469104d3f6
relation.isAuthorOfPublication.latestForDiscovery76fdc4d1-7624-4332-a9d0-f06826000679
sigepi.autor.atividadeJOSEMARY ANGELICA CORREA GONCALVES:924:240:S
sigepi.autor.atividadePATRICIA DE LARA ANTONIO:6585:330:N
sigepi.autor.atividadeLINDA V. E. CALDAS:1495:330:N
sigepi.autor.atividadeCARMEN CECILIA BUENO TOBIAS:1592:240:N

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
31146.pdf
Tamanho:
444.32 KB
Formato:
Adobe Portable Document Format

Licença do Pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:

Coleções