Practical leak detection and leak rate measurement applied to a hot cell for radiopharmaceutical preparations in a GMP-compliant laboratory

dc.contributor.authorSANTOS, R. S.
dc.contributor.authorZAPPAROLI, C. L.
dc.contributor.authorSANTOS, J. M.
dc.contributor.authorCAETANO, G. F. R.
dc.contributor.authorCAVALCANTI, M. G. P.
dc.contributor.authorANDRADE, D. A.
dc.coverageInternacional
dc.date.accessioned2026-05-15T15:23:37Z
dc.date.available2026-05-15T15:23:37Z
dc.date.issued2026
dc.description.abstractHot cells are containment systems used in radiopharmaceutical production, designed to ensure radiation protection and product quality. Leak tightness testing is a mandatory requirement under CNEN NN 6.13 standard, as it directly impacts worker safety, environmental protection, and product integrity. This study aimed to perform a practical pressure change leak test in a hot cell used for radiopharmaceutical preparations, in accordance with applicable technical standards. The pressure change method was applied using pressures between 1.6 and 2.5 times the working pressure for a testing time of 15 minutes. The method proved practical, requiring no highly specialized equipment, although its sensitivity to temperature highlights the importance of a high-precision thermometer. Results showed that while the method is suitable for ISO 10648-2:1994 Class 2 compliance, it may not be adequate for Class 1 due to the risk of mathematically inaccurate results under temperature variations. Multiple rounds of leak detection and repair were required to achieve the required containment level. Overall, the DC, RIC, CIC, and Solid Waste chambers met Class 2 leak tightness requirements, being suitable for use in radiopharmacy facilities.
dc.format.extent1-17
dc.identifier.citationSANTOS, R. S.; ZAPPAROLI, C. L.; SANTOS, J. M.; CAETANO, G. F. R.; CAVALCANTI, M. G. P.; ANDRADE, D. A. Practical leak detection and leak rate measurement applied to a hot cell for radiopharmaceutical preparations in a GMP-compliant laboratory. <b>Brazilian Journal of Radiation Sciences</b>, v. 14, n. 1, p. 1-17, 2026. DOI: <a href="https://dx.doi.org/10.15392/2319-0612.2026.3004">10.15392/2319-0612.2026.3004</a>. Disponível em: https://repositorio.ipen.br/handle/123456789/49877.
dc.identifier.doi10.15392/2319-0612.2026.3004
dc.identifier.fasciculo1
dc.identifier.issn2319-0612
dc.identifier.orcidhttps://orcid.org/0000-0002-6689-3011
dc.identifier.percentilfiSem Percentil F.I.
dc.identifier.percentilfiCiteScoreSem Percentil CiteScore
dc.identifier.urihttps://repositorio.ipen.br/handle/123456789/49877
dc.identifier.vol14
dc.language.isoeng
dc.relation.ispartofBrazilian Journal of Radiation Sciences
dc.rightsopenAccess
dc.titlePractical leak detection and leak rate measurement applied to a hot cell for radiopharmaceutical preparations in a GMP-compliant laboratory
dc.title.alternativeDetecção de vazamentos e medição da taxa de vazamento em célula quente para produção de radiofármacos em um laboratório BPF
dc.typeArtigo de periódico
dspace.entity.typePublication
ipen.autorRODRIGO DA SILVEIRA SANTOS
ipen.autorCARLOS LEONEL ZAPPAROLI JUNIOR
ipen.autorJOEL MENDES DOS SANTOS
ipen.autorDELVONEI ALVES DE ANDRADE
ipen.codigoautor15558
ipen.codigoautor424
ipen.codigoautor16000
ipen.codigoautor1258
ipen.contributor.ipenauthorRODRIGO DA SILVEIRA SANTOS
ipen.contributor.ipenauthorCARLOS LEONEL ZAPPAROLI JUNIOR
ipen.contributor.ipenauthorJOEL MENDES DOS SANTOS
ipen.contributor.ipenauthorDELVONEI ALVES DE ANDRADE
ipen.identifier.fiSem F.I.
ipen.identifier.fiCiteScoreSem CiteScore
ipen.identifier.ipendoc31964
ipen.type.genreArtigo
relation.isAuthorOfPublication36919fe6-e18b-479b-becd-411c07b54914
relation.isAuthorOfPublicationa07fd69d-5af1-4a66-8326-65b71ef03a19
relation.isAuthorOfPublication6b953641-f2f0-4dcd-b641-7b2dd7b219dd
relation.isAuthorOfPublication0eeb4436-68e5-4573-a603-35f5fc912178
relation.isAuthorOfPublication.latestForDiscovery36919fe6-e18b-479b-becd-411c07b54914
sigepi.autor.atividadeRODRIGO DA SILVEIRA SANTOS:15558:420:N
sigepi.autor.atividadeCARLOS LEONEL ZAPPAROLI JUNIOR:424:420:N
sigepi.autor.atividadeJOEL MENDES DOS SANTOS:16000:-1:N
sigepi.autor.atividadeDELVONEI ALVES DE ANDRADE:1258:420:N

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
31964.pdf
Tamanho:
1.07 MB
Formato:
Adobe Portable Document Format

Licença do Pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:

Coleções