Micro-homogeneity evaluation of a bovine kidney candidate reference material

Carregando...
Imagem de Miniatura
Data
2019
Data de publicação:
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Brazilian Journal of Radiation Sciences
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
The minimum sample intake for which a reference material remains homogeneous is one of the parameters that must be estimated in the homogeneity assessment study of reference materials. In this work, Instrumental Neutron Activation Analysis was used to evaluate this quantity in a bovine kidney candidate reference material. The mass fractions of 9 inorganic constituents were determined in subsamples between 1 and 2 mg in order to estimate the relative homogeneity factor (HE) and the minimum sample mass to achieve 5% and 10% precision on a 95% confidence level. Results ob-tained for HE in all the analyzed elements were satisfactory. The estimated minimum sample intake was between 2 mg and 40 mg, depending on the element.

Como referenciar
CASTRO, L.; MOREIRA, E.G.; VASCONCELLOS, M.B.A. Micro-homogeneity evaluation of a bovine kidney candidate reference material. Brazilian Journal of Radiation Sciences, v. 7, n. 2A, p. 1-9, 2019. DOI: 10.15392/bjrs.v7i2A.576. Disponível em: http://repositorio.ipen.br/handle/123456789/30000. Acesso em: 13 Feb 2025.