Efficiency stability of HPGe detectors under distinct count rates

dc.contributor.authorZAHN, G.S.pt_BR
dc.contributor.authorGENEZINI, F.A.pt_BR
dc.coverageNacionalpt_BR
dc.date.accessioned2021-08-03T13:51:05Z
dc.date.available2021-08-03T13:51:05Z
dc.date.issued2021pt_BR
dc.description.abstractIn this work the reproducibility of the efficiency of different HPGe detectors was analyzed under a series of different conditions. The detectors studied were plugged either to a regular analogical amplifier or to a digital signal processing (DSP) device, to evaluate the possible differences between either setup. Detectors were inspected by performing a long series of sequential measurements with standard calibration sources and comparing the standard deviation of the number of counts per second in each series to the uncertainty of the individual measurements. Detectors were also subjected to distinct count rates, to verify the possible experimental issues associated with this parameter. The results allow a discussion on the stability of the detectors’ efficiencies over a few days, the possible dependence with the count rate, and the estimation of the uncertainty related to the efficiency variation.pt_BR
dc.format.extent1-8pt_BR
dc.identifier.citationZAHN, G.S.; GENEZINI, F.A. Efficiency stability of HPGe detectors under distinct count rates. <b>Brazilian Journal of Radiation Sciences</b>, v. 9, n. 1A, p. 1-8, 2021. DOI: <a href="https://dx.doi.org/10.15392/bjrs.v9i1A.1410">10.15392/bjrs.v9i1A.1410</a>. Disponível em: http://repositorio.ipen.br/handle/123456789/32110.
dc.identifier.doi10.15392/bjrs.v9i1A.1410pt_BR
dc.identifier.fasciculo1Apt_BR
dc.identifier.issn2319-0612pt_BR
dc.identifier.orcid0000-0002-6318-6805pt_BR
dc.identifier.orcid0000-0003-3237-8588pt_BR
dc.identifier.orcidhttps://orcid.org/0000-0002-6318-6805
dc.identifier.orcidhttps://orcid.org/0000-0003-3237-8588
dc.identifier.percentilfiSem Percentilpt_BR
dc.identifier.urihttp://repositorio.ipen.br/handle/123456789/32110
dc.identifier.vol9pt_BR
dc.relation.ispartofBrazilian Journal of Radiation Sciencespt_BR
dc.rightsopenAccesspt_BR
dc.sourceMeeting on Nuclear Applications (ENAN), 14th, October 21-25, 2019, Santos, SPpt_BR
dc.subjectcalibration standards
dc.subjectcomparative evaluations
dc.subjectcounting ratemeters
dc.subjectdata covariances
dc.subjectefficiency
dc.subjecthigh-purity ge detectors
dc.subjectstability
dc.titleEfficiency stability of HPGe detectors under distinct count ratespt_BR
dc.typeArtigo de periódicopt_BR
dspace.entity.typePublication
ipen.autorFREDERICO ANTONIO GENEZINI
ipen.autorGUILHERME SOARES ZAHN
ipen.codigoautor2045
ipen.codigoautor950
ipen.contributor.ipenauthorFREDERICO ANTONIO GENEZINI
ipen.contributor.ipenauthorGUILHERME SOARES ZAHN
ipen.date.recebimento21-08
ipen.identifier.fiSem F.I.pt_BR
ipen.identifier.ipendoc27881pt_BR
ipen.type.genreArtigo
relation.isAuthorOfPublication0c41d307-45a9-47c4-8281-5aa9ed46e6a4
relation.isAuthorOfPublicationde99ffee-d830-4bea-aeeb-e8f3ccc34921
relation.isAuthorOfPublication.latestForDiscoveryde99ffee-d830-4bea-aeeb-e8f3ccc34921
sigepi.autor.atividadeGENEZINI, F.A.:2045:310:Npt_BR
sigepi.autor.atividadeZAHN, G.S.:950:310:Spt_BR

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
27881.pdf
Tamanho:
758.96 KB
Formato:
Adobe Portable Document Format
Descrição:

Licença do Pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:

Coleções