Efficiency stability of HPGe detectors under distinct count rates

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Brazilian Journal of Radiation Sciences
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
In this work the reproducibility of the efficiency of different HPGe detectors was analyzed under a series of different conditions. The detectors studied were plugged either to a regular analogical amplifier or to a digital signal processing (DSP) device, to evaluate the possible differences between either setup. Detectors were inspected by performing a long series of sequential measurements with standard calibration sources and comparing the standard deviation of the number of counts per second in each series to the uncertainty of the individual measurements. Detectors were also subjected to distinct count rates, to verify the possible experimental issues associated with this parameter. The results allow a discussion on the stability of the detectors’ efficiencies over a few days, the possible dependence with the count rate, and the estimation of the uncertainty related to the efficiency variation.

Como referenciar
ZAHN, G.S.; GENEZINI, F.A. Efficiency stability of HPGe detectors under distinct count rates. Brazilian Journal of Radiation Sciences, v. 9, n. 1A, p. 1-8, 2021. DOI: 10.15392/bjrs.v9i1A.1410. Disponível em: http://repositorio.ipen.br/handle/123456789/32110. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento

Coleções