In this work, a theoretical-experimental methodology was developed for the calibration of HPGe Detector for quantitative analysis of iodine-125. Photopick efficiencies was measured between 0 - 40 keV using a certified calibration source (iodine-129) with well-known activity. The efficiency curve was determined for an energy range that covers the region of interest for iodine-125 (0 - 40 keV). The calibrate spectrum was obtained for study and the result was saved in the detector directory for future analysis and determination of the iodine-125 activity.
Como referenciar
CORREIA, RUANYTO W.; ROSTELATO, MARIA E.C.M.; ZEITUNI, CARLOS A. Study and calibration of the HPGe detector for radionuclide analysis of iodine-125. In: INTERNATIONAL NUCLEAR ATLANTIC CONFERENCE, October 21-25, 2019, Santos, SP. Proceedings... Rio de Janeiro: Associação Brasileira de Energia Nuclear, 2019. p. 370-382. Disponível em: http://repositorio.ipen.br/handle/123456789/30360. Acesso em: 14 Mar 2025.