Activimeter “in situ” calibration methodology to 111In and 123I

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Brazilian Journal of Radiation Sciences
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
The activimeter calibration has the purpose of ensure greater reliability in measurement results, hence the activimeters used are commonly installed in controlled areas and, in some cases, with difficult access. The activimeter “in situ” calibration methodology presented in this work allows its execution only with the displacement of the radioactive samples and not of the activimeter itself, which simplifies the procedure of nuclear medicine services and at radiopharmaceuticals production centers, without affecting the quality and accuracy of measurements. After the application of the methodology by qualified technicians, the obtained results of the tested activimeters showed its importance since the calibration factors can present correction of up to 5% for 111In and greater than 5% for 123I.

Como referenciar
MARTINS, E.W.; POTIENS, M.P.A. Activimeter “in situ” calibration methodology to 111In and 123I. Brazilian Journal of Radiation Sciences, v. 11, n. 1A, p. 1-9, 2023. DOI: 10.15392/2319-0612.2023.2212. Disponível em: http://repositorio.ipen.br/handle/123456789/34215. Acesso em: 20 Mar 2026.