Activimeter “in situ” calibration methodology to 111In and 123I

dc.contributor.authorMARTINS, E.W.pt_BR
dc.contributor.authorPOTIENS, M.P.A.pt_BR
dc.coverageNacional
dc.date.accessioned2023-11-24T21:48:42Z
dc.date.available2023-11-24T21:48:42Z
dc.date.issued2023pt_BR
dc.description.abstractThe activimeter calibration has the purpose of ensure greater reliability in measurement results, hence the activimeters used are commonly installed in controlled areas and, in some cases, with difficult access. The activimeter “in situ” calibration methodology presented in this work allows its execution only with the displacement of the radioactive samples and not of the activimeter itself, which simplifies the procedure of nuclear medicine services and at radiopharmaceuticals production centers, without affecting the quality and accuracy of measurements. After the application of the methodology by qualified technicians, the obtained results of the tested activimeters showed its importance since the calibration factors can present correction of up to 5% for 111In and greater than 5% for 123I.pt_BR
dc.format.extent1-9pt_BR
dc.identifier.citationMARTINS, E.W.; POTIENS, M.P.A. Activimeter “in situ” calibration methodology to 111In and 123I. <b>Brazilian Journal of Radiation Sciences</b>, v. 11, n. 1A, p. 1-9, 2023. DOI: <a href="https://dx.doi.org/10.15392/2319-0612.2023.2212">10.15392/2319-0612.2023.2212</a>. Disponível em: http://repositorio.ipen.br/handle/123456789/34215.
dc.identifier.doi10.15392/2319-0612.2023.2212pt_BR
dc.identifier.fasciculo1Apt_BR
dc.identifier.issn2319-0612
dc.identifier.orcid0000-0002-4049-6720pt_BR
dc.identifier.orcidhttps://orcid.org/0000-0002-4049-6720
dc.identifier.percentilfiSem Percentil
dc.identifier.percentilfiCiteScoreSem Percentil CiteScore
dc.identifier.urihttp://repositorio.ipen.br/handle/123456789/34215
dc.identifier.vol11pt_BR
dc.relation.ispartofBrazilian Journal of Radiation Sciences
dc.rightsopenAccesspt_BR
dc.sourceINTERNATIONAL JOINT CONFERENCE RADIO, 6th, August 15-19, 2022, Poços de Caldas, MG.pt_BR
dc.subjectcalibration
dc.subjectactivity meters
dc.subjectmeasuring methods
dc.subjectin-situ processing
dc.titleActivimeter “in situ” calibration methodology to 111In and 123Ipt_BR
dc.typeArtigo de periódicopt_BR
dspace.entity.typePublication
ipen.autorMARIA DA PENHA ALBUQUERQUE POTIENS
ipen.autorELAINE WIRNEY MARTINS
ipen.codigoautor346
ipen.codigoautor5968
ipen.contributor.ipenauthorMARIA DA PENHA ALBUQUERQUE POTIENS
ipen.contributor.ipenauthorELAINE WIRNEY MARTINS
ipen.date.recebimento23-11
ipen.identifier.fiSem F.I.
ipen.identifier.fiCiteScoreSem CiteScore
ipen.identifier.ipendoc29844
ipen.type.genreArtigo
relation.isAuthorOfPublication30f213ae-321d-4d51-8763-0a2585255fb9
relation.isAuthorOfPublicationa0291c92-b89b-4010-87cb-a17fe0b08289
relation.isAuthorOfPublication.latestForDiscoverya0291c92-b89b-4010-87cb-a17fe0b08289
sigepi.autor.atividadePOTIENS, M.P.A.:346:330:Npt_BR
sigepi.autor.atividadeMARTINS, E.W.:5968:330:Spt_BR

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
29844.pdf
Tamanho:
799.72 KB
Formato:
Adobe Portable Document Format
Descrição:

Licença do Pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:

Coleções