Evaluation of a special parallel-plate ionization chamber for mammography and digital breast tomosynthesis dosimetry

dc.contributor.authorSOARES, CLIMERIO S.
dc.contributor.authorCALDAS, LINDA V.E.
dc.coverageInternacional
dc.creator.eventoSIMPOSIO REPROLAM, 1st
dc.date.accessioned2025-01-20T19:50:43Z
dc.date.available2025-01-20T19:50:43Z
dc.date.eventoNovember 5-8, 2024
dc.description.abstractA special parallel-plate ionization chamber, developed at the Calibration Laboratory of IPEN (LCI), was tested to verify the possibility of its application in mammography and digital breast tomosynthesis dosimetry. It has a rectangular box geometric shape (146 mm x 46.4 mm x 24 mm), a sensitive length of 10 cm, and a sensitive volume of 3.2 cm3. The chamber wall is made of PMMA, its collecting electrode is made of graphite, and its chamber window is a Mylar foil. The chamber was submitted to several characterization and quality control tests: leakage current, short- and medium-term stabilities, stabilization time, saturation, ion collection efficiency, polarity effect, linearity of response, and angular dependence. The maximum value observed in the leakage current test was 0.53% so within the recommended limit of ±5% by the IEC 61674 standard. In the short-term stability test, the highest coefficient of variation obtained was 0.17%. This value satisfies the IEC 61674 standard requirements (±3%). The maximum value observed in the medium-term stability was below the recommended limit of ±2%. For the stabilization time test, the difference between the ionization currents obtained between 15 min and 60 min after switching on the measuring system was only 0.62%. This result is within the recommended limits of ±2% of response variation, according to the IEC 61674 standard requirements. In the saturation test, it was achieved in the whole voltage interval evaluated. For all pairs of voltages tested during the saturation determination, the polarity effect was within the recommended limit of ±1%, as stated by IEC 61674. The ion collection efficiency, in turn, was better than 99.99%. The special ionization chamber presented a linear behavior, with a regression coefficient of R2 = 0.9999. In the angular test, the maximum variation obtained was only 1.7%, so below the 3% recommended by the IEC 61674 standard. Results will also be presented about the energy dependence of the ionization chamber response in standard direct and attenuated mammography beams.
dc.description.sponsorshipConselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)
dc.description.sponsorshipFundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
dc.description.sponsorshipIDCNPq: 403664/2022-5; 305142/2021-6; 406303/2022-3
dc.description.sponsorshipIDFAPESP: 18/05982-0
dc.identifier.citationSOARES, CLIMERIO S.; CALDAS, LINDA V.E. Evaluation of a special parallel-plate ionization chamber for mammography and digital breast tomosynthesis dosimetry. In: SIMPOSIO REPROLAM, 1st, November 5-8, 2024, Recife, PE. <b>Abstract...</b> Disponível em: https://repositorio.ipen.br/handle/123456789/48879.
dc.identifier.orcidhttps://orcid.org/0000-0002-7362-2455
dc.identifier.urihttps://repositorio.ipen.br/handle/123456789/48879
dc.local.eventoRecife, PE
dc.publisherRed de Optimización de la Protección Radiológica en LatinoAmerica y Caribe
dc.rightsopenAccess
dc.titleEvaluation of a special parallel-plate ionization chamber for mammography and digital breast tomosynthesis dosimetry
dc.typeResumo de eventos científicos
dspace.entity.typePublication
ipen.autorCLIMERIO SANTOS SOARES
ipen.autorLINDA V. E. CALDAS
ipen.codigoautor15306
ipen.codigoautor1495
ipen.contributor.ipenauthorCLIMERIO SANTOS SOARES
ipen.contributor.ipenauthorLINDA V. E. CALDAS
ipen.event.datapadronizada2024
ipen.identifier.ipendoc30936
ipen.notas.internasAbstract
ipen.type.genreResumo
relation.isAuthorOfPublicationcf41e896-7e2b-454d-9ac2-a68a692374c2
relation.isAuthorOfPublication7f46d4f4-dfd6-4485-a767-10df5b4f4f13
relation.isAuthorOfPublication.latestForDiscoverycf41e896-7e2b-454d-9ac2-a68a692374c2
sigepi.autor.atividadeCLIMERIO SANTOS SOARES:15306:330:S
sigepi.autor.atividadeLINDA V. E. CALDAS:1495:330:N

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
30936.pdf
Tamanho:
497.64 KB
Formato:
Adobe Portable Document Format

Licença do Pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição: