Calibration of surface radioactive contamination detectors with beta and alpha radiation

Carregando...
Imagem de Miniatura

Data

Data de publicação

2006

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

É parte de

INTERNATIONAL SYMPOSIUM ON RADIATION PHYSICS, 10th
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
The Calibration Laboratory of Instituto de Pesquisas Energéticas e Nucleares, IPEN, has already over 35 years been calibrating instruments used in radiation protection and therapy measurements that belongs to hospitals, industries, clinics and other users located in São Paulo state and in other parts of Brazil. At the present time, the Calibration Laboratory acts in the Radiation Protection, Radiation Therapy, Nuclear Medicine and Diagnostic Radiology areas, using special set-ups with gamma and beta radiation sealed sources, alpha and beta radiation plain sources and low and intermediate energies of X radiation. The instruments used in radiation protection measurements represent 88% of the tested instruments by the laboratory annually (approximately 1900). About 190 instruments are used in nuclear medicine clinics to detect surface radioactive contamination. The objective of this work is to present the results about this kind of calibration using alpha and beta radiation. The instruments normally used for surface contamination measurements are of pancake type Geiger tubes with very thin mica windows (2 mg.cm-2). They usually are sensitive to beta particles, with energies greater than 40 keV. They can also detect energetic alpha particles and gamma-rays. However, in this case the efficiency will be very different depending on the different radionuclides. The counting efficiency also depends on how close to the contaminated surface the detector is held. Surface contamination is usually expressed in units of radioactivity per area unit (Bq/m2 ) [1]. Other units such as picocuries per 100 cm2 or disintegrations per minute per square centimetre ae sometimes used. Special set-ups of plain beta (90Sr + 90Y, 99Tc, 36Cl and 137Cs) and alpha (241Am) sources were utilized. The surface emission rates varied from 406 to 1300 s-1, with an uncertainty of 5%. The instruments efficiencies obtained were between 0.02 to 0.45% for alpha radiation and 0.12 to 0.58 to beta radiation.

Como referenciar
POTIENS, M.P.A.; COSTA, A.M.; CARVALHO, V.S.; CALDAS, L.V.E. Calibration of surface radioactive contamination detectors with beta and alpha radiation. In: INTERNATIONAL SYMPOSIUM ON RADIATION PHYSICS, 10th, September 17-22, 2006, Coimbra, Portugal. Abstract... Disponível em: http://repositorio.ipen.br/handle/123456789/20406. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento