Study of the gamma radiation response of watch glasses

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Radiation Measurements
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
Some dosimetric properties of watch glasses were studied applying the thermoluminescence technique. The watch glass samples were powdered, and the selected grains were mixed with TeflonTM. The mixture was pressed and sintered to produce pellets of watch glass-TeflonTM composites. The glow curves of the pellets show two peaks at 130 and 195 {ring operator} C. Reproducibility of TL response was estimated to have a maximum coefficient of variation of 4.0%. The dose-response curve is sublinear between 0.5 and 20.0 kGy. The calibration curve is linear between 1.0 Gy and 1.0 kGy. The minimum detection limits were also determined. The gamma radiation dose response and the thermal stability of the materials were studied with the purpose to establish the best conditions of watch glasses for use in gamma radiation dosimetry.

Como referenciar
TEIXEIRA, MARIA I.; COSTA, ZELIA M. da; COSTA, CELIA R. da; PONTUSCHKA, WALTER M.; CALDAS, LINDA V.E. Study of the gamma radiation response of watch glasses. Radiation Measurements, v. 43, p. 480-482, 2008. DOI: 10.1016/j.radmeas.2007.11.029. Disponível em: http://repositorio.ipen.br/handle/123456789/5061. Acesso em: 20 Mar 2026.