Performance comparison of an extrapolation chamber in computed tomography standard beams of two laboratories

dc.contributor.authorCASTRO, MAYSA C.pt_BR
dc.contributor.authorSILVA, NATALIA F.pt_BR
dc.contributor.authorSANTOS, WILLIAM S.pt_BR
dc.contributor.authorCALDAS, LINDA V.E.pt_BR
dc.coverageInternacionalpt_BR
dc.date.accessioned2021-12-10T15:35:32Z
dc.date.available2021-12-10T15:35:32Z
dc.date.issued2021pt_BR
dc.description.abstractThe doses received by the patients during the computed tomography (CT) exams are the highest when compared to other diagnostic radiology procedures. This fact makes it so important to keep the quality control using the correct instrument. For dosimetry in CT, a pencil type ionization chamber is usually utilized with a sensitive volume length of 10 cm. A comparison of the performance of an extrapolation chamber in two different laboratories, at the Instrument Calibration Laboratory (LCI) from Instituto de Pesquisas Energéticas e Nucleares (IPEN, Brazil) and at the National Physical Laboratory (NPL, UK) was undertaken. Two chamber characteristics were determined (zero depth and extrapolation curve), and the main recommended characterization tests were performed (stabilization time, saturation curve, polarity effect, ion collection efficiency, linearity of response, and variation of response with source-detector distance). This extrapolation chamber was calibrated using the standard system of NPL laboratory, and the energy dependence was determined. The results obtained were within the international recommendations of IEC 61674, except for the energy dependence, as already expected. This study also established a reference system for the LCI of the IPEN, using a homemade extrapolation chamber for X-ray CT beams. Using the extrapolation curves obtained in the characterization tests and the correction factors of ionic recombination, and scattering and fluorescence, the air kerma rates were obtained, as a reference system in standard laboratory computed tomography beams. The last correction factor was obtained using computer simulations by the MCNP5 code (Monte Carlo). Moreover, two pencil type ionization chambers were calibrated using the replacement method in relation to the extrapolation chamber.pt_BR
dc.description.sponsorshipConselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq)pt_BR
dc.description.sponsorshipFundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)pt_BR
dc.description.sponsorshipIDCNPq: 301335/2016-8pt_BR
dc.description.sponsorshipIDFAPESP: 18/05982-0pt_BR
dc.format.extent1-8pt_BR
dc.identifier.citationCASTRO, MAYSA C.; SILVA, NATALIA F.; SANTOS, WILLIAM S.; CALDAS, LINDA V.E. Performance comparison of an extrapolation chamber in computed tomography standard beams of two laboratories. <b>Radiation Physics and Chemistry</b>, v. 189, p. 1-8, 2021. DOI: <a href="https://dx.doi.org/10.1016/j.radphyschem.2021.109748">10.1016/j.radphyschem.2021.109748</a>. Disponível em: http://repositorio.ipen.br/handle/123456789/32385.
dc.identifier.doi10.1016/j.radphyschem.2021.109748pt_BR
dc.identifier.issn0969-806Xpt_BR
dc.identifier.orcid0000-0002-7362-2455pt_BR
dc.identifier.orcidhttps://orcid.org/0000-0002-7362-2455
dc.identifier.percentilfi57.09pt_BR
dc.identifier.percentilfiCiteScore70.00pt_BR
dc.identifier.urihttp://repositorio.ipen.br/handle/123456789/32385
dc.identifier.vol189pt_BR
dc.relation.ispartofRadiation Physics and Chemistrypt_BR
dc.rightsopenAccesspt_BR
dc.subjectcomputerized tomography
dc.subjectionization chambers
dc.subjectextrapolation chambers
dc.subjectcalibration
dc.subjectmonte carlo method
dc.subjectsimulation
dc.titlePerformance comparison of an extrapolation chamber in computed tomography standard beams of two laboratoriespt_BR
dc.typeArtigo de periódicopt_BR
dspace.entity.typePublication
ipen.autorLINDA V. E. CALDAS
ipen.autorNATALIA FIORINI DA SILVA
ipen.autorMAYSA COSTA DE CASTRO
ipen.codigoautor1495
ipen.codigoautor12139
ipen.codigoautor12105
ipen.contributor.ipenauthorLINDA V. E. CALDAS
ipen.contributor.ipenauthorNATALIA FIORINI DA SILVA
ipen.contributor.ipenauthorMAYSA COSTA DE CASTRO
ipen.date.recebimento21-12
ipen.identifier.fi2.776pt_BR
ipen.identifier.fiCiteScore4.5pt_BR
ipen.identifier.ipendoc28153pt_BR
ipen.identifier.iwosWoSpt_BR
ipen.range.fi1.500 - 2.999
ipen.range.percentilfi50.00 - 74.99
ipen.type.genreArtigo
relation.isAuthorOfPublication7f46d4f4-dfd6-4485-a767-10df5b4f4f13
relation.isAuthorOfPublicationc8a4b1ff-5955-47b9-a15a-ecbd12a4baa8
relation.isAuthorOfPublicationa7ec93af-7e13-4572-8409-2fcafeab7015
relation.isAuthorOfPublication.latestForDiscoverya7ec93af-7e13-4572-8409-2fcafeab7015
sigepi.autor.atividadeCALDAS, LINDA V.E.:1495:330:Npt_BR
sigepi.autor.atividadeSILVA, NATALIA F.:12139:330:Npt_BR
sigepi.autor.atividadeCASTRO, MAYSA C.:12105:330:Spt_BR

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
28153.pdf
Tamanho:
3.29 MB
Formato:
Adobe Portable Document Format
Descrição:

Licença do Pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
license.txt
Tamanho:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descrição:

Coleções