Comparative study of spectrophotometric response of the 270 bloom fricke gel dosimeter to clinical photon and electron beams
Carregando...
Data
Data de publicação
2010
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
INTERNATIONAL SYMPOSIUM ON STANDARDS, APPLICATIONS AND QUALITY ASSURANCE IN MEDICAL RADIATION DOSIMETRY
Resumo
This study aims to compare the spectrophotometric response of the Fricke xylenol gel (FXG) dosimeter developed at IPEN, prepared using 270 Bloom gelatine from porcine skin produced in Brazil, for clinical photons 6 to 18 MV and electron beams with energies between 6 to 16 MeV, to the reference depth using liquid and virtual water phantoms. The optical absorption spectra, dosimetric wavelengths, energy and dose dependent response, sensitivity and lower detection limits were evaluated and compared for both clinical beams. All results obtained in this study are satisfactory and indicate the viability of implementing this dosimeter in photon and electron 3D dosimetry.
Como referenciar
CAVINATO, C.C.; SAKURABA, R.K.; CRUZ, J.C.; CAMPOS, L.L. Comparative study of spectrophotometric response of the 270 bloom fricke gel dosimeter to clinical photon and electron beams. In: INTERNATIONAL SYMPOSIUM ON STANDARDS, APPLICATIONS AND QUALITY ASSURANCE IN MEDICAL RADIATION DOSIMETRY, November 9-12, 2010, Vienna, Austria. Proceedings... Disponível em: http://repositorio.ipen.br/handle/123456789/17844. Acesso em: 30 Dec 2025.