Development and tests of a 30cmpencil-type ionization chamber for dosimetry in standard and clinical CT beams

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Biomedical Physics & Engineering Express
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
The use of computed tomography (CT) for imaging procedures is growing due to advances in the CT equipment technology. To perform the dosimetry inCT beams, the most widely used instrument is the pencil type ionization chamber with a sensitive volume length of 100 mm; however, some studies have shown that this dosimeter has underestimated the dose values. Therefore, in this study an ionization chamber with sensitive volume length of 300mmwas developed at IPEN. The characterization of this chamber was performed, and the results were obtained within the international recommended limits. As an application, the developed ionization chamber was tested in clinical beams of two different hospitals.

Como referenciar
CASTRO, M.C.; SILVA, N.F.; XAVIER, M.; ANDRADE, M.E.A.; KHOURY, H.J.; CALDAS, L.V.E. Development and tests of a 30cmpencil-type ionization chamber for dosimetry in standard and clinical CT beams. Biomedical Physics & Engineering Express, v. 3, n. 5, p. 1-7, 2017. DOI: 10.1088/2057-1976/aa82e3. Disponível em: http://repositorio.ipen.br/handle/123456789/28457. Acesso em: 30 Dec 2025.