New core configuration for the fabrication of 125I radioactive sources for cancer treatment

Carregando...
Imagem de Miniatura
Data
2020
Data de publicação:
Autores IPEN
ANSELMO FEHER
LUCAS VERDI ANGELOCCI
MARIA ELISA CHUERY MARTINS ROSTELATO
OSVALDO LUIZ DA COSTA
JOÃO AUGUSTO MOURA
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Applied Radiation and Isotopes
ODS
ODS 3
Exportar
Mendeley
Projetos de Pesquisa
Unidades Organizacionais
Fascículo
Resumo
In order to provide prostate brachytherapy treatment for more Brazilian men, IPEN is building a laboratory for the manufacture of radioactive sources. The new methodology for the production of iodine-125 seeds with yield 71.7% ± 5.3%. Points of importance were evaluated/discussed: photo-sensibility, reaction vial type, the substitution for iodine-131, pH, and solution volume. The surface was analyzed by FTIR and EDS. At the end, a Monte Carlo-MCNP6 simulation was performed to evaluate the TG-43 parameters.

Como referenciar
SOUZA, CARLA D. de; ZEITUNI, CARLOS A.; FEHER, ANSELMO; MOURA, JOÃO A.; COSTA, OSVALDO L. da; ANGELOCCI, LUCAS V.; ROSTELATO, MARIA E.C.M. New core configuration for the fabrication of 125I radioactive sources for cancer treatment. Applied Radiation and Isotopes, v. 165, p. 1-8, 2020. DOI: 10.1016/j.apradiso.2020.109307. Disponível em: http://200.136.52.105/handle/123456789/31590. Acesso em: 15 Jun 2024.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento