Response characteristics of a tandem ionization chamber in standard X-ray beams
Carregando...
Data
Data de publicação
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Applied Radiation and Isotopes
Resumo
A tandem ionization chamber was developed for quality control programs of the kilovoltage X-ray equipment used in radiotherapy. This chamber was tested and calibrated in low-energy X-ray beams in accordance with international recommendations, and it showed a satisfactory level of performance, mainly with regard to its main use, namely formation of a tandem system for confirmation of half-value layers in X-radiation beams previously determined by the conventional method. In this developed system no absorbers or special setups are necessary.
Como referenciar
COSTA, A.M.; CALDAS, L.V.E. Response characteristics of a tandem ionization chamber in standard X-ray beams. Applied Radiation and Isotopes, v. 58, n. 4, p. 495-500, 2003. DOI: 10.1016/S0969-8043(03)00026-5. Disponível em: http://repositorio.ipen.br/handle/123456789/5887. Acesso em: 30 Dec 2025.