Autoclaving and battery capacity influence on laser fluorescence measurements

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Acta Odontologica Scandinavica
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
Aim. To evaluate the influence of probe tip autoclaving and depleted alkaline batteries on laser fluorescence (LF-DIAGNOdent) device performance. Material and methods. One-hundred-and-twenty occlusal sites were analyzed with an LF device in nine conditions: No autoclaved probe and probe after 5, 10, 15, 20, 25, 30, 40, and 50 autoclaving cycles. Subsequently, the performance of the device was analyzed with: New batteries, batteries with 1.49/1.39 V, 1.38/1.37 V, 1.36/1.34 V, 1.33/1.32 V, and lower than 1.32 V. LF values, sensitivity, specificity, accuracy, and area under the ROC curve were compared. Results. In the probe study, sensitivity was lower after 50 sterilizing cycles, though specificity was higher than the assessment performed using a new tip. In the batteries study, specificity was higher for depleted batteries, but LF performance did not differ significantly among the groups. Conclusion. Batteries do not significantly influence device performance, but consecutive sterilization of probes in autoclave alters readings, downgrading its performance.

Como referenciar
CABRAL, RENATA M.R.; MENDES, FAUSTO M.; MIURA, FABIANA; RIBEIRO, ADRIANA da C.; BRAGA, MARIANA M.; ZEZELL, DENISE M. Autoclaving and battery capacity influence on laser fluorescence measurements. Acta Odontologica Scandinavica, v. 66, p. 122-127, 2008. DOI: 10.1080/00016350802017484. Disponível em: http://repositorio.ipen.br/handle/123456789/5010. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento

Coleções