TANDEM study with thermoluminescent materials in standard diagnostic beams
Carregando...
Data
Data de publicação
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Revista Brasileira de Pesquisa e Desenvolvimento
Resumo
Tandem systems formed with different
thermoluminescent dosemeters were evaluated
in standard diagnostic radiation beams. The
utilized materials are commercially available,
and the aim of this study was to carry out a
performance comparison. The results show that
the energy dependence of the materials, except
TLD-100, were very high. Therefore, there are
several possibilities of pair combinations forming
a tandem system, but all of them using TLD-100
with a high energy dependent TL material. For
all pair combinations evaluated in this study,
the differences in the measurement ratios of
subsequent energy values were always higher
than 14%. The main differences among these
tandem systems were in the dose sensitivity.
The more advantageous TLD pair, among all
evaluated, was TLD-100 and TLD-200, but several
other combinations were also very efficient.