TANDEM study with thermoluminescent materials in standard diagnostic beams

dc.contributor.authorMAIA, ANA F.pt_BR
dc.contributor.authorCALDAS, LINDA V.E.pt_BR
dc.coverageNacionalpt_BR
dc.date.accessioned2014-07-15T13:44:50Zpt_BR
dc.date.accessioned2014-07-30T11:48:06Z
dc.date.available2014-07-15T13:44:50Zpt_BR
dc.date.available2014-07-30T11:48:06Z
dc.date.issued2007pt_BR
dc.description.abstractTandem systems formed with different thermoluminescent dosemeters were evaluated in standard diagnostic radiation beams. The utilized materials are commercially available, and the aim of this study was to carry out a performance comparison. The results show that the energy dependence of the materials, except TLD-100, were very high. Therefore, there are several possibilities of pair combinations forming a tandem system, but all of them using TLD-100 with a high energy dependent TL material. For all pair combinations evaluated in this study, the differences in the measurement ratios of subsequent energy values were always higher than 14%. The main differences among these tandem systems were in the dose sensitivity. The more advantageous TLD pair, among all evaluated, was TLD-100 and TLD-200, but several other combinations were also very efficient.
dc.format.extent97-100pt_BR
dc.identifier.citationMAIA, ANA F.; CALDAS, LINDA V.E. TANDEM study with thermoluminescent materials in standard diagnostic beams. <b>Revista Brasileira de Pesquisa e Desenvolvimento</b>, v. 9, n. 3, p. 97-100, 2007. Disponível em: http://repositorio.ipen.br/handle/123456789/5156.
dc.identifier.fasciculo3pt_BR
dc.identifier.issn0104-7698pt_BR
dc.identifier.orcidhttps://orcid.org/0000-0002-7362-2455
dc.identifier.urihttp://repositorio.ipen.br/handle/123456789/5156pt_BR
dc.identifier.vol9pt_BR
dc.relation.ispartofRevista Brasileira de Pesquisa e Desenvolvimentopt_BR
dc.rightsopenAccessen
dc.subjectbeamspt_BR
dc.subjectdosimetrypt_BR
dc.subjectluminescent dosemeterspt_BR
dc.subjectphosphoruspt_BR
dc.subjectradiation protectionpt_BR
dc.subjectthermoluminescent dosemeterspt_BR
dc.subjectthermoluminescent dosimetrypt_BR
dc.titleTANDEM study with thermoluminescent materials in standard diagnostic beamspt_BR
dc.typeArtigo de periódicopt_BR
dspace.entity.typePublication
ipen.autorANA FIGUEIREDO MAIA
ipen.autorLINDA V. E. CALDAS
ipen.codigoautor2991
ipen.codigoautor1495
ipen.contributor.ipenauthorANA FIGUEIREDO MAIA
ipen.contributor.ipenauthorLINDA V. E. CALDAS
ipen.date.recebimento08-09pt_BR
ipen.identifier.fiSem F.I.pt_BR
ipen.identifier.ipendoc12815pt_BR
ipen.type.genreArtigo
relation.isAuthorOfPublicationde6ea8e6-1554-4d6d-b73a-40e3df506cd9
relation.isAuthorOfPublication7f46d4f4-dfd6-4485-a767-10df5b4f4f13
relation.isAuthorOfPublication.latestForDiscovery7f46d4f4-dfd6-4485-a767-10df5b4f4f13
sigepi.autor.atividadeMAIA, ANA F.:2991:-1:Spt_BR
sigepi.autor.atividadeCALDAS, LINDA V.E.:1495:28:Npt_BR

Pacote Original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
12815.pdf
Tamanho:
3.13 MB
Formato:
Adobe Portable Document Format
Descrição:

Coleções