Preliminary results on energy dependence of the fricke gel dosimeter for low energy x-radiation
Carregando...
Data
Data de publicação
2008
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
INTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATION, 12th
Resumo
The conventional Fricke dosimeter have been commonly used for calibration of various procedures
that use ionizing radiation and the Fricke gel dosimeter, which consist of the modification of conventional
dosimeter by the addition of gelatinous agent, have been studied for radiotherapy application. The present study
was carried out with the purpose of to investigate the energy dependence of the Fricke gel dosimeter response for
low energy X-radiation, in order to know the properties of this dosimeter for future application other than
radiotherapy. The Fricke gel dosimeters were irradiated using a PANTAK SEIFERT X-rays systems model
ISOVOLT 160 HS with absorbed dose of 1 Gy with different X-rays beam qualities. The Fricke gel response was
evaluated by optical absorption spectrophotometry technique. The absorbance values correspond to the
wavelength of 550 nm. The energy dependence of the Fricke gel dosimeter response for low energy X-radiation
relative to 60Co corrected to cuvettes wall absorption increases between 0.107 ± 0.162 to 7.38 keV until
1.000 ± 0.192 to 30.48 keV.
Como referenciar
CAVINATO, CHRISTIANNE C.; CAMPOS, LETICIA L. Preliminary results on energy dependence of the fricke gel dosimeter for low energy x-radiation. In: INTERNATIONAL CONGRESS OF THE INTERNATIONAL RADIATION PROTECTION ASSOCIATION, 12th, October 19-24, 2008, Buenos Aires, Argentina. Proceedings... Disponível em: http://repositorio.ipen.br/handle/123456789/13185. Acesso em: 20 Mar 2026.