Influence of the vials radioactive residue in Nuclear Medicine procedure applied to a new “in situ” activimeter calibration methodology

Carregando...
Imagem de Miniatura

Data

Data de publicação

2017

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

É parte de

INTERNATIONAL TOPICAL MEETING ON INDUSTRIAL RADIATION AND RADIOISOTOPE MEASUREMENT APPLICATIONS, 10th
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
A safe and efficient Nuclear Medicine Services (NMS) performance depends, among other factors, on a complete quality assurance program, mainly in the case of the activity meters. Many elements may influence this equipment response, being the major errors related to vials characteristics, such as volume and geometry, as presented in other studies [1,2]. Correction factors must be applied in order to avoid patient exposure to excessive doses.However, another aspect that should be taken into account is how much radioactive material is lost during the radioisotope labeling, a process that includes many recipient changes.In this study we present the estimated activity lost during 99mTc labeling process and show how important is to add this value to the total activity measured in a new “in situ” calibration procedure, in which the activity meter is not removed from its place of use. For the radioactive residue determination99mTc labeling procedure was reproduced under laboratorial conditions, resulting in losses of up to 40 %. However greater errors are expected in a NMS. Activity meter calibration using two methodologies: the first consists in a direct comparison between the two systems with the same vial (P6-P6 and 10R-10R), and the second is a comparison made using two different vials (P6-10R), with and without geometry and residual activity corrections. A calibration factor of 0.999 was obtained in the reference situation. The same procedure performed without the application of the necessary correction factors resulted in a variation of almost 3 %, which is high considering the test was made under controlled conditions and less residue was left inside vials. The “in situ” calibration procedure worked well and was important to present the need of correction factors application, including the residue activity measurement.

Como referenciar
KUAHARA, LILIAN T.; CORREA, EDUARDO; POTIENS, MARIA da P.A. Influence of the vials radioactive residue in Nuclear Medicine procedure applied to a new “in situ” activimeter calibration methodology. In: INTERNATIONAL TOPICAL MEETING ON INDUSTRIAL RADIATION AND RADIOISOTOPE MEASUREMENT APPLICATIONS, 10th, July 09-13, 2017, Chicago, IL, USA. Abstract... p. 65-65. Disponível em: http://repositorio.ipen.br/handle/123456789/27951. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.

Agência de fomento