Improvement of an exoelectron emission detection system for radiation dosimetry
Carregando...
Data
Data de publicação
2024
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
SIMPOSIO REPROLAM, 1st
Resumo
Solid-state dosimeters are fundamental components for the areas of radiation metrology, medical radiology, radiotherapy and
environmental monitoring. This work explores the promising potential of the phenomenon of exoelectron emission (EE) for
applications in these areas. Through an innovative system employing a continuous flow gas proportional detector, a precise
and reliable method was developed to measure dose in materials like beryllium oxide (BeO), carbon-doped alumina
(Al2O3:C), and calcium sulfate (CaSO4), commonly used in dosimetry. The meticulously designed and constructed system
comprises a continuous flow gas proportional detector coupled to a complex electronic circuit. This circuit, incorporating an
Amptek charge-sensitive preamplifier and an amplifier and pulse shaper, was carefully calibrated to detect the EE signals
emitted by the dosimeters following a radiation exposure. BeO, Al2O3:C, and CaSO4 dosimeters, in the form of 5 mm
diameter pellets, were exposed to beta (90Sr/90Y) and alpha (241Am) radiations at varying doses and heated in an electric
oven at controlled temperatures. The dosimeter EE signal was measured by the developed system and meticulously analyzed.
The results obtained were promising. The heating system exhibited remarkable stability and consistency throughout the
measurement process, ensuring the reproducibility of the measurements. The amplifier circuit designed to detect exoelectrons
yielded consistent and accurate results for the exposure of the tested dosimeters, demonstrating their ability to measure the
absorbed dose. A key advantage of the system lies in its use of a proportional detector, eliminating the need for expensive
photomultiplier tube and equipment. Moreover, the developed instrumentation exhibits notable stability over time,
minimizing the need for frequent calibrations. This work presents a novel and promising system as an alternative for solidstate
dosimetry, utilizing the EE technique and a continuous flow gas proportional detector. The system stands out for its
precision, reproducibility, low cost, and stability, making it a valuable tool for applications in radiation metrology and
calibrations.
Como referenciar
SILVEIRA, IURY S.; GOMES, FRANCISCO F.; CALDAS, LINDA V.E. Improvement of an exoelectron emission detection system for radiation dosimetry. In: SIMPOSIO REPROLAM, 1st, November 5-8, 2024, Recife, PE. Abstract... Disponível em: https://repositorio.ipen.br/handle/123456789/48882. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.