Quality control tests for calibration of KAP metes
Carregando...
Data
Data de publicação
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Brazilian Journal of Radiation Sciences
Resumo
In accordance with IEC 60580:2021, KAP meters are typically calibrated using cylindrical reference ionization chambers aligned with RQR radiation qualities specified by ISO 61267. However, discrepancies in manufacturing geometry between these reference chambers and PKA meters prompted a series of quality control tests to establish an effective calibration methodology. Two distinct tests of quality control were conducted to compare the performance of a reference KAP meter (PDC - Radcal®) with a reference ionization chamber (Radcal® 10x5-180), using the Agfa® PANTAK Constant Potential X-Ray System. These tests included homogeneity assessments with varying horizontal displacements of the reference PDC meter and reproducibility tests at consistent positions. The two tests were conducted for Kerma and PKA unities. A third test is still in progress, the long-term stability, with repeated measurements; as with the tests ago, for Kerma and KAP measurements. These results demonstrated good correlation and traceability between the PDC and the ionization chamber, with reproducibility and homogeneity within an uncertainty margin below 3 %. These findings validate the proposed calibration method for practical applications.
Como referenciar
ALMEIDA JUNIOR, J.N.; POTIENS, M.P.A.; RODRIGUES JUNIOR, O. Quality control tests for calibration of KAP metes. Brazilian Journal of Radiation Sciences, v. 12, n. 4B, p. 1-12, 2024. DOI: 10.15392/2319-0612.2024.2687. Disponível em: https://repositorio.ipen.br/handle/123456789/49613. Acesso em: 29 Apr 2026.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.