Insights and simulation of metascintillator-based brain PET systems to enhance ToF capability
Carregando...
Data
Data de publicação
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Nuclear Engineering and Technology
Resumo
The integration of Time-of-Flight (ToF) information into Positron Emission Tomography (PET) image reconstruction
enhances both signal-to-noise ratio and the localization of annihilation events. A critical component
contributing to the accuracy of ToF-PET is the scintillator. To overcome the time resolution limitations
in conventional scintillators, the metascintillator approach has been proposed. The metascintillator is an
engineered composition of small units that combines and optimizes various features within a single scintillator
heterostructure.
In this work, metascintillator-based brain PET systems were simulated using the GATE toolkit and compared
with designs based on bulk LYSO or BGO. Sensitivity, noise equivalent count rate (NECR) and scatter fraction
were evaluated following NEMA guidelines. To match the peak sensitivity of a system utilizing a 15 mm bulk
BGO, the metascintillator-based scanners using BGO/BaF2, BGO/EJ232, LYSO/BaF2 and LYSO/EJ232 must
possess thicknesses of 23.2 mm, 22.5 mm, 29.7 mm and 31.1 mm, respectively. With ToF gain, the scanner
utilizing a 25 mm thick LYSO-EJ232 metascintillator exhibited the most promising NECR curve, peaking at
1180 cps at 1600 MBq. This work takes a significant step towards harnessing the information gain facilitated
by the integration of metascintillator-based detectors in PET imaging.
Como referenciar
BONIFACIO, D.A.B.; LATELLA, R.; MURATA, H.M.; BENLLOCH, J.M.; GONZALEZ, A.J.; LECOQ, P.; KONSTANTINOU, G. Insights and simulation of metascintillator-based brain PET systems to enhance ToF capability. Nuclear Engineering and Technology, v. 57, n. 5, p. 1-10, 2024. DOI: 10.1016/j.net.2024.11.037. Disponível em: https://repositorio.ipen.br/handle/123456789/49830. Acesso em: 27 Apr 2026.