Dose-rate constant and air-kerma strength evaluation of a new 125I brachytherapy source using Monte-Carlo
Carregando...
Data
Data de publicação
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Brazilian Journal of Radiation Sciences
Resumo
Brachytherapy is a modality of radiotherapy which treats tumors using ionizing radiation with sources located close to the tumor. The sources can be produced from several radionuclides in various formats, such as Iodine-125 seeds and Iridium-192 wires. In order to produce a new Iodine-125 seed in IPEN/CNEN and ensure its quality, it is essential to describe the seed dosimetry, so when applied in a treatment the lowest possible dose to neighboring healthy tissues can be reached. The report by the AAPM’s Task Group 43 U1 is a document that indicates the dosimetry procedures in brachytherapy based on physical and geometrical parameters. In this study, dose-rate constant and air-kerma strength parameters were simulated using the Monte Carlo radiation transport code MCNP4C. The air-kerma strength is obtained from an ideal modeled seed, since its actual value should be measured for seeds individually in a specialized lab with a Wide-Angle Free-Air Chamber (WAFAC). Dose-rate constant and air-kerma strength are parameters that depends on intrinsic characteristics of the source, i.e. geometry, radionuclide, encapsulation, and together they define the dose-rate to the reference point. Radial dose function describes the dose fall-off with distance from the source. This study presents the values found for these parameters with associated statistical uncertainty, and is part of a larger project that aims the full dosimetry of this new seed model, including experimental measures.
Como referenciar
PRIMO, C.O.; ANGELOCCI, L.V.; KARAM JUNIOR, D.; ZEITUNI, C.A.; ROSTELATO, M.E.C.M. Dose-rate constant and air-kerma strength evaluation of a new 125I brachytherapy source using Monte-Carlo. Brazilian Journal of Radiation Sciences, v. 9, n. 1A, p. 1-11, 2021. DOI: 10.15392/bjrs.v9i1A.1441. Disponível em: http://repositorio.ipen.br/handle/123456789/32116. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.