Radiosurgery dosimetry using CaSO4:Eu OSLD film
Carregando...
Data
Data de publicação
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Radiation Protection Dosimetry
Resumo
Recent studies demonstrated that optically stimulated luminescence (OSL) systems allow the evaluation of doses for 2D mapping in a relatively fast and simple way and results show submillimeter resolution. This work presents, for the first time, an optically stimulated luminescence dosemeter (OSLD) in the form of film made with CaSO4:Eu particles embedded in a silicone elastomer matrix. The OSLD film was produced using a low-cost and relatively simple methodology. This film is reusable and the signal can be satisfactorily bleached using blue LEDs. The main dosimetric properties were evaluated using TL/OSL Risø reader with blue stimulation and Hoya U-340 filter. Investigation shows repeatability within 3% when measuring with the same film sample. Regarding the OSLD film homogeneity, nearly 12% sensitivity change was observed within the 5 × 5 cm2 produced film. Additionally, the dose response curve shows linearity from 5 to 25 Gy. Fading of the OSL signal is relatively high, about 50% in the first week and then is stable. Nevertheless, a 3 × 3 cm2 OSLD film was successfully used to map dose distribution in radiosurgery (6 MV photon beam). This work demonstrates the feasibility of 2D dosimetry using reusable OSLD films based on CaSO4:Eu.
Como referenciar
GASPARIAN, PATRICIA B.R.; MALTHEZ, ANNA L.M.C.; TORQUATO, MICHELE; CAMPOS, LETICIA L. Radiosurgery dosimetry using CaSO4:Eu OSLD film: a feasibility study. Radiation Protection Dosimetry, v. 199, n. 8-9, p. 1040-1046, 2023. DOI: 10.1093/rpd/ncad109. Disponível em: https://repositorio.ipen.br/handle/123456789/48001. Acesso em: 30 Dec 2025.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.