Comparison of calibration of surface contamination monitors with Brazilian Network - 2018/2019

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Journal of Physics: Conference Series
ODS
ODS 7ODS 9ODS 12
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
This work reports the results obtained from the comparison involving 7 laboratories in Brazil. This exercise is about the calibration service of surface contamination monitors. The monitor has been calibrated to BS ISO 7503-3, the calibration factor in terms of surface emission rate. The comparison was conducted by the Brazilian National Laboratory of Ionizing Radiation Metrology (LNMRI / IRD) from October 2018 to July 2019. The extensive sources used were 14C, 137Cs, 60Co, 90Sr/90Y, 36Cl and 241Am. The result of this proficiency test was excellent and proved the calibration capacity of the Brazilian network in the calibration service for surface contamination monitors.

Como referenciar
CABRAL, T.S.; LARANJEIRA, A.S.; POTIENS, M.P.A.; SOARES, C.M.A.; KHOURY, H.; SAITO, V.; DAVID, M.; VIDAL, L.C.; PIRES, M.A. Comparison of calibration of surface contamination monitors with Brazilian Network - 2018/2019. Journal of Physics: Conference Series, v. 1826, p. 1-6, 2021. DOI: 10.1088/1742-6596/1826/1/012039. Disponível em: http://repositorio.ipen.br/handle/123456789/32032. Acesso em: 30 Dec 2025.