A rotational calibration method using thermoluminescent dosemeters for dose determination in computed tomography beams

Carregando...
Imagem de Miniatura

Data

Data de publicação

Orientador

Título da Revista

ISSN da Revista

Título do Volume

É parte de

É parte de

É parte de

Nuclear Instruments and Methods in Physics Research, B
Exportar
Mendeley

Projetos de Pesquisa

Unidades Organizacionais

Fascículo

Resumo
A procedure for calibration of thermoluminescent dosimeters, called rotational calibration, was developed to create a procedure more adequate for dose procedures used in computed tomography. Thermoluminescent dosimeters were rotated during irradiation to simulate the set-up used in dosimetric procedures in computed tomography equipment. Three commercial types of thermoluminescent dosimeters were calibrated using this methodology. The results showed that the measured values were lower when the dosimeters were irradiated with rotation than in a static geometry. Although the reading differences were not very large, they are not negligible, and they contribute to underestimating the dose.

Como referenciar
MAIA, ANA F.; CALDAS, LINDA V.E. A rotational calibration method using thermoluminescent dosemeters for dose determination in computed tomography beams. Nuclear Instruments and Methods in Physics Research, B, v. 266, n. 1, p. 107-110, 2008. DOI: 10.1016/j.nimb.2007.10.008. Disponível em: http://repositorio.ipen.br/handle/123456789/5106. Acesso em: 20 Mar 2026.
Esta referência é gerada automaticamente de acordo com as normas do estilo IPEN/SP (ABNT NBR 6023) e recomenda-se uma verificação final e ajustes caso necessário.