Dosimetric evaluation and comparation of TL responses of LiF:Mg,Ti and μLiF:Mg,Ti in the clinical electron beams dosimetry applied to total skin irradiation (TSEB) treatments
Carregando...
Data
Data de publicação
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Radiation Measurements
ODS
Resumo
The Total Skin Electron Beam (TSEB) irradiation is a radiotherapeutic technique that aims to provide the patient's
skin surface with a more homogeneous dose, in order to treat cutaneous T-cell lymphomas, both for
curative and palliative purposes. Electron irradiation penetrates a few millimeters into the skin, reaching the
affected parts completely, without penetrating the internal organs. In vivo dosimetry has become an important
role for the treatment of total skin irradiation within a rigorous quality assurance program that should be an
integral part of the radiotherapy departments. The use of TLDs in vivo can identify variations in the prescribed
dose because its measurement accuracy and great precision. The LiF:Mg,Ti is the most used TL material and
widely studied in radiotherapy dosimetry due to near tissue-equivalence of the material, along with its overall
reliability. The dosimeters of μLiF:Mg,Ti have been gaining considerable importance in the radiotherapy departments.
These detectors allow measurements in vivo with great advantages due to their minimum dimensions
of 1×1×1 mm3. This paper reports a comparative study of the TL responses of both materials to dose evaluation
in TSEB treatments. The TL response of both materials in several TSEB parameter tests and in clinical
application were evaluated, analyzing the dose distribution in a treatment simulation using AldersonRando
anthropomorphic phantom. The results showed that the μLiF:Mg,Ti presented greater variation of the response
in relation to LiF dosemeters in some parameters analyzed, due to the small dimensions and to evaluate doses
absorbed in the surface over a large area in the treatment plan.
Como referenciar
ALMEIDA, S.B.; VILLANI, D.; SAKURABA, R.K.; REZENDE, A.C.P.; CAMPOS, L.L. Dosimetric evaluation and comparation of TL responses of LiF:Mg,Ti and μLiF:Mg,Ti in the clinical electron beams dosimetry applied to total skin irradiation (TSEB) treatments. Radiation Measurements, v. 125, p. 15-18, 2019. DOI: 10.1016/j.radmeas.2019.03.007. Disponível em: http://repositorio.ipen.br/handle/123456789/29975. Acesso em: 24 Mar 2026.