Energy dependent response of the Fricke gel dosimeter prepared with 270 bloom gelatine for photons in the energy range 13.93 KeV-6MeV
Carregando...
Data
Data de publicação
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Nuclear Instruments and Methods in Physics Research, A
Resumo
The spectrophotometric energy dependent response to photons with effective energies between
13.93 keV and 6 MeV of the Fricke xylenol gel (FXG) dosimeter developed at IPEN, prepared using
270 Bloom gelatine, was evaluated in order to verify the possible dosimeter application in other
medicine areas in addition to radiosurgery, for example, breast radiotherapy and blood bags
radiosterilization. Other dosimetric characteristics were also evaluated. The obtained results indicate
that the FXG dosimeter can contribute to dosimetry in different medical application areas including
magnetic resonance imaging (MRI) evaluation technique that permits three-dimensional (3D) dose
distribution evaluation.
Como referenciar
CAVINATO, C.C.; CAMPOS, L.L. Energy dependent response of the Fricke gel dosimeter prepared with 270 bloom gelatine for photons in the energy range 13.93 KeV-6MeV. Nuclear Instruments and Methods in Physics Research, A, v. 619, n. 1-3, p. 198-202, 2010. DOI: 10.1016/j.nima.2009.10.144. Disponível em: http://repositorio.ipen.br/handle/123456789/4693. Acesso em: 20 Mar 2026.