X-ray spectrometry applied for determination of linear attenuation coefficient of tissue-equivalent materials
Carregando...
Data
Data de publicação
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Radiation Physics and Chemistry
Resumo
Resin-based materials equivalent to water were developed and four different samples were obtained. The linear
attenuation coefficients of all samples was evaluated using X-ray spectrometry with primary and transmitted
beams using voltages at the ranging from 60 to 120 kV. The experimental measured values were compared with
theoretical reference values to water and with that obtained using the Least Square Method algorithm methodology
(method applied to diagnostic radiology and radiotherapy). Our results show that differences between
the measured values and the target μ(E) was lower than 7% ± 0.3 in the energy range from 20 to 80 keV. These
results enable to consider that the material developed and produced is a good option to be used as a waterequivalent
material and the experimental method adequate to its quantitative evaluation.
Como referenciar
FRIMAIO, AUDREW; NASCIMENTO, BRUNA C.; BARRIO, RAMON M.M.; CAMPOS, LETICIA L.; COSTA, PAULO R. X-ray spectrometry applied for determination of linear attenuation coefficient of tissue-equivalent materials. Radiation Physics and Chemistry, v. 160, p. 89-95, 2019. DOI: 10.1016/j.radphyschem.2019.03.021. Disponível em: http://repositorio.ipen.br/handle/123456789/30057. Acesso em: 30 Dec 2025.