Phantom validation of optical attenuation coefficient estimation model with depth resolution for Optical Coherence Tomography
Carregando...
Data
Data de publicação:
2019
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
É parte de
ENCONTRO DE OUTONO DA SOCIEDADE BRASILEIRA DE FÍSICA, 42.
Resumo
Optical Coherence Tomography (OCT) has been used for imaging and analyzes over a broad number of
applications.[1-3] Besides its use for morphological analysis [3] the OCT signal has been used to determine the
optical attenuation coe±cient of biological samples for classi¯cation and diagnostic proposes. [1,2] The most
common model for estimating the optical attenuation coe±cient based on OCT signal relies on the Lambert-
Beers law. It usually assumes a constant attenuation coe±cient value over the image range or over a selected
range, losing the in-depth resolution on OCT. Previous published work on literature developed a model to esti-
mate the attenuation coe±cient with depth resolution [4], however it assumes that the light is totally attenuated
within the image depth range failing for membrane like samples. We present a model that, using the tissue
sample transmittance as input, remove this limitation and to estimate the depth-resolved optical attenuation
coe±cient. This method allows us to obtain an image of tissue optical properties instead of that from intensity
contrast, guiding diagnosis and tissues di®erentiation, extending its application from thin to tick samples. The
performance of our method was tested with the assistance of a home built single layer and multi-layer phantoms
( 100¹m each layer). These optical phantoms are composed of a substrate polydimethylsiloxane (PDMS), Zinc-
Phthalocyanine (ZnPc) dye as chromophores, and TiO2 as scattering agent. The optical attenuation coe±cient
ranges from 0:9 to 2:32 mm¡1, measured using an integrating sphere followed by the Inverse Adding Doubling
processing technique. We show that the estimated depth-resolved attenuation coe±cient recovers the reference
values, with a error deviation of 7 %.
Como referenciar
AMARAL, MARCELLO M.; ARAUJO, JEANN C.R.; ANTUNES, ANDREA; MONTE, ADAMO F.G.; CARA, ANA C.B. de; FREITAS, ANDERSON Z. de; ZEZELL, DENISE M. Phantom validation of optical attenuation coefficient estimation model with depth resolution for Optical Coherence Tomography. In: ENCONTRO DE OUTONO DA SOCIEDADE BRASILEIRA DE FÍSICA, 42., 26-31 de maio, 2019, Aracaju, SE. Resumo... São Paulo: Sociedade Brasileira de Física, 2019. Disponível em: http://repositorio.ipen.br/handle/123456789/30217. Acesso em: 16 Mar 2025.