Intrinsic efficiency of CaSO4:Dy to electron beams
Carregando...
Data
Data de publicação
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Radioproteção
Resumo
The intrinsic efficiency of CaSO4
:Dy thermoluminescent dosimeters,
developed by Instituto de Pesquisas Energéticas e Nucleares – IPEN was calculated
for 3.43, 5.48, 8.27, 11.67 and 15.42 MeV clinical electron beams at their depth
of maximum ionization in polymethyl methacrylate (PMMA), SW Solid Water
and standard water phantoms. The sensitivity, varying from 14.6 ± 0.3 mC.Gy-1
(11.67 MeV, PMMA) to 19.6 ± 0.5 mC.Gy-1 (15.42 MeV, water), the reproducibility
of the TL response, better than 2.4%, and the TL response as a function of the
dose, linear between 0.01 and 3.25 Gy, were also evaluated. The intrinsic efficiency
of 0.0342 ± 0.0021 % indicates that CaSO4
:Dy dosimeters can be applied to the
radiation protection for high energy electron beams generated by linear accelerators.