Experimental Validation of the LabSOCS Detector Efficiency Simulation
Carregando...
Data
Data de publicação
Autores IPEN
Orientador
Título da Revista
ISSN da Revista
Título do Volume
É parte de
É parte de
É parte de
Brazilian Journal of Radiation Sciences
Resumo
The precise knowledge of the detection efficiency for gamma ray spectrometers
is often of paramount importance, and its experimental determination can be both timeconsuming
and challenging, especially for complex geometries and/or extensive sources.
A common solution for that is the use of Monte Carlo simulations, and some companies
have developed commercial solutions. In the present work, the accuracy of the efficiency
values determined by Mirion Industries’ LabSOCS detection efficiency simulator was
assessed by determining the activities of point sources measured under five distinct
geometries, and comparing the results to the certified activity values. The results show
that, while the software delivers reasonably reliable results, it tends to overestimate the
efficiency, and special care may have to be taken with the precision of the geometrical
measurements.
Como referenciar
ZAHN, G.S.; GENEZINI, F.A. Experimental Validation of the LabSOCS Detector Efficiency Simulation. Brazilian Journal of Radiation Sciences, v. 12, p. 01-11, 2024. 4A. DOI: 10.15392/2319-0612.2024.2641. Disponível em: https://repositorio.ipen.br/handle/123456789/49826. Acesso em: 29 Apr 2026.