Resumo
Single-particle inductively coupled plasma-mass spectrometry (sp-ICP-MS) is a powerful
technique that enables the simultaneous measurement of nanoparticle size and the quantification
of metal-containing nanoparticles at real concentrations. These measurements are critical for
understanding the potential uses of nanoparticles in various fields. However, sp-ICP-MS is not
yet considered a mature methodology. Standardization is necessary, particularly in assessing
the reliability of size distribution measurements. This study aims to standardize sp-ICP-MS by
assessing the reliability of size distribution measurements for polyvinylpyrrolidone-coated silver
nanoparticles and quantifying the input quantities contributing to uncertainties. The uncertainties
in calibration, repeatability, and trueness were evaluated based on the thresholds defined by
International Organization for Standardization (ISO) standards. Calibration uncertainty was found
to be 8.1%, while repeatability was 0.04%, both within the stipulated range of less than 10%.
The expanded uncertainty was calculated to be 17%, with a 95% coverage probability for the
reference material. Limits of detection (LOD) and quantification (LOQ) for dissolved concentration,
particles per mL, and size were determined to be 0.37 μg L–1, 97.5 particles mL–1, and 24.6 nm,
respectively. These results demonstrate that the reliability and repeatability of sp-ICP-MS meet
ISO-defined thresholds, suggesting that with further standardization, sp-ICP-MS could become a
reliable methodology for nanoparticle analysis.